Construction of Moving-Iron Power Factor Meter
Construction of Moving-Iron Power Factor Meter
Points : Construction of Moving-Iron Power Factor Meter
One type of power factor meter suitable for 3-phase balanced
circuits is shown. It consists of three fixed coils R, Y and B with
axis mutually at 120° and intersecting on the centre line of the
instrument, these coils are connected respectively in R, Y and B
lines of the 3-phase supply through current transformers. When so
energised, the three coils produce a synchronously rotating flux.
There is a fixed coil B at the centre of three fixed coils and is
connected in series with a high resistance across one of the pair of
lines, say, across R and Y lines as shown. Coil B is threaded by the
instrument spindle which carries an iron cylinder C to which are
fixed sector-shaped iron vanes V1 and V2. The
same spindle also carries damping vanes and pointer (not shown in
fig.) but there are no control springs.